9.3 Adjusting SS sensor offset
You must adjust the SS sensor offset when activating or changing the SS sensor. This operation should be
carried out only once.
Use the
and
keys to select
the configuration required.
Select the
Sensor offset adjustment
menu and press
ENTER.
Cover the optical slot with a dark coloured ruler and press
ENTER.
CONFIG
WHAT DO YOU WANT TO DO?
ESC: Escape menu
Averaging sensor response
Sludge slope coefficient reminder
Sensor offset adjustment
Graphic screen
4-20mA outputs adjustment
Relay adjustment
Measure units
Language
Factory default value reminder
CONFIG
EXTINCTION OF THE I.R. LIGHT
BEAM
Occult the optical slot
with an opaque ruler
Sensor offset: 12mV
DO NOT VALIDATE
Extinction incomplete!
ESC: Escape menu
Enter: Validate and record
If the offset is too high check that the dark coloured
ruler is positioned properly over the optical slot
(see following page)
No
Yes
CONFIG
WHAT DO YOU WANT TO DO?
Averaging sensor response
Sludge slope coefficient reminder
Sensor offset adjustment
Graphic screen
4-20mA outputs adjustment
Relay adjustment
Measure units
Language
Factory default value reminder
ESC: Escape menu
Offset Correct