6.1.6 Adjusting the temperature sensor slope:
If you observe a small measurement error, you may only adjust the slope of your sensor by carrying out the
following procedure:
Use the
and
keys to select
the type of calibration and the
probe to be calibrated.
Select the
Gain adjust
menu of the temperature sensor and
press ENTER.
Use the
and
keys to adjust the
ambient water temperature value.
Press the ENTER key to confirm slope
adjustment.
CALIBRATION
DU TEMPERATURE SENSOR
Immerse the sensor in
water at ambient temperature
Standard:
Measure:
:
°C
°C
20.00
20.20
WAIT: Nonstable measurement
AWAIT MEASUREMENT STABILISATION
Immerse the sensor in water at ambient temperature.
ESC: Leave the current slope
ENTER: Validate the new slope
ENTER : Validate the choice
CALIBRATION
SENSORS CALIBRATION
SS SENSOR
Gain adjust
Immediate calibration
Differed calibration
Theoretical calibration
TEMPERATURE SENSOR
Gain adjust
Complete calibration
Theoretical calibration
ENTER: Validate the choice