User manual
PONSEL
Référence : NOTICE_ACTEON_2050_v001
V : 002
26/67
CALIBRATION
OF SS SENSOR
ESC: Cancel the procedure
ENTER: Validate the calibration
Use the
and
keys to adjust
sludge concentration.
Content of SS after analysis
Lab measure: 5.00g/l
Recorded measure!
4.80
g/l
Current slope coefficient: 100%
New slope coefficient: 120%
Comment: The measurement
window is displayed again
once the calibration has been
correctly carried out.
20.00
°C
-24 hrs
5.00
MEASURE
g/l
INFO
CONF
HIST
CAL
LOG
MEAS
Mode1 R1
R2
ACTEON 2050-20 Turbidimeter
g/l
1
6
If the second calibration point is incorrect, an error
message window appears (see "Error message
information" chapter §6.2.2)
Calibration
Correct
No
Yes
Press the ENTER key to confirm the second
calibration point