6.2 SS sensor calibration error message
6.2.1 CLEAN WATER calibration error
One of two messages may be displayed depending on the calibration error.
1) First type of information:
Your SS sensor response is too low:
1)
Clean the sensor head with a water jet, focussing on the insides of the optical measuring channel.
2)
Replace the water with clean water.
2) Second type of information:
Your SS sensor response is too high:
1)
Check the water temperature.
If you are unable to calibrate the SS sensor after these checks, seek advice from an after-sales service (see last
page).
6.2.2 SLUDGE calibration error
Your SS sensor response is too low:
CALIBRATION
Out of range! Calibration impossible
Check the sensor or purity of water:
1)
The sensor could be clogged
2)
Water could include bubbles or
become turbid
If the checking is negative, check the
sensor.
ESC: Escape menu
SS SENSOR
CALIBRATION IMPOSSIBLE
CALIBRATION
Out of range! Calibration impossible
Check the sensor:
If the water temperature is <10°C or
>35°C, await the measurement
stabilisation.
Otherwise, check the sensor.
ESC: Escape menu
OF SS SENSOR
CALIBRATION IMPOSSIBLE
CALIBRATION
Out of range! Calibration impossible
Check the sensor and standard sludge:
Possibility of incorrect SS titration
(factor >10) or optics clogged.
If the checking is negative, check the
sensor.
ESC: Escape menu
SS SENSOR
CALIBRATION IMPOSSIBLE