POM2-3333
Merger Unit Test User
Manual
7
2.2 Technical specifications
Current generator
Setting
Outputting current amplifier, frequency and phase of each
phase are adjustable independently
Amplitude
6×12.5A/phase
Accuracy
±0.05%(0.2A
~
12.5A),<0.1mA(0.05A
~
0.2A)
Resolution
1mA
Pmax
6×12.5A/the phase time
,≥80VA/
phase
;
3×25A/phase,
≥160VA/
phase
Current rise/fall time
<100µs
THD%
≤0.05%(0.2A
~
12.5A)
Frequency
1Hz
~
1kHz
amplitude-frequency
characteristic
Amplitude changes
≤±0.05% (45Hz
~
65Hz)
≤±0.2%(10Hz
~
45Hz ,65Hz
~
450Hz)
≤±0.5% (450Hz
~
1000Hz)
Output time
<
5A
/
phase, continuous output
;
5A
~
12.5A
/
phase,
>
180s
Abnormal condition alarm
Overload, distortion, open circuit detection and alarm
automatically
Voltage generator
Setting
Outputting vulgate amplifier, frequency and phases pf each
phase are adjustable independently
Amplitude
6×120V/phase
Degree of accuracy
±0.05%(4V
~
120V),<2mV(0.5V
~
4V)
Resolving power
1 mV(0.2V
~
10V),10 mV(10V
~
120V)
Maximum output power
≥60VA/
phase, 3×120V/the phase time
≥30VA/
phase, 6×120V/the phase time
vulgate rise/fall time
<100µs
THD%
≤0.05% (10V
~
120V)
Frequency
1Hz
~
1kHz
amplitude-frequency
characteristic
Amplitude changes
≤±0.05% (45Hz
~
65Hz)
≤±0.2%(10Hz
~
45Hz ,65Hz
~
450Hz)
≤±0.5% (450Hz
~
1000Hz)
Output time
Under the condition of rated , continuous output
Components Security
Overload, distortion, short circuit detection and alarm
automatically
Synchronous error between
current and vulgate
≤10µs
General
Frequency
Sinusoidal signal
1Hz
~
1000Hz
Degree of accuracy
<
0.5mHz (1Hz
~
65Hz)
<
5mHz(65 Hz
~
450Hz)
<
10mHz(450 Hz
~
1000Hz)
Resolution ratio
0.0001Hz
Output characteristic
Can superimpose random amplifier of harmonic and direct
current 2 to 13 times.
Phase
Phase angle range
0 to 359. 9°
Degree of accuracy
±0.05°
resolution ratio
0.01°