PF3000 Power Filter Characteristic Tester User Manual
software will remind overheat fault.
2.4. Environment Conditions and Factors
The basic working conditions are for factory inspection and arbitral test. The rated working
conditions are for product adjustment and operation.
Factor
Basic condition
Rated condition
Temperature
20
℃±2℃
-5
℃
~
+45
℃
Relative humidity
45%
~
75%
10%
~
90%
Atmospheric pressure
86kPa
~
106kPa
80kPa
~
110kPa
AC
power
supply
Voltage
220V±1%
220V±10%
Frequency
49.5Hz
~
50.5Hz
49Hz
~
51Hz
Wave
Sine wave
THD≤2%
Sine wave
THD≤5%
2.5. Safety Performance
Indoor temperature and humidity <75%, the insulation resistance meets:
1000V megger for power supply
(metal chasis), resistance≥300MΩ
500V megger for voltage output (metal chasis), resistance≥50MΩ
2.6. Shell Protection
Shell protection standard applies with the IP21 regulated in GB 4208-1993.
9