7
III. Instructions
3.1 First use
➢
After being unpacked, the medical refrigerator should be placed well to according to the service
environment requirement;
➢
Although the refrigerator has been cleaned inside when leaving factory, it is advised to scrub the medical
refrigerator with warm water added with a little neutral detergent and then scrub it with clean water and
wipe it up (electrical system cannot be cleaned but scrubbed with dry cloth);
➢
Set the temperature controller of the medical refrigerator at 5
℃
and turn on the power. After 3 minutes,
the compressor is started. After 30 minutes, the temperature inside the medical refrigerator decreases
obviously. It means the refrigerating system is operating smoothly and the machine test is finished;
➢
For first use, it is advised to put the goods to be stored in the refrigerator after the inside temperature falls
to the operating temperature. If there are too much goods to be stored, it is advised to store separately
for three times. When the inside temperature of the refrigerator falls to the set temperature after the
storage of the last time, store the goods for the second time. Storing too much goods at one time may
cause that the refrigerator cannot fall to the set temperature for a long time, thereby causing damages of
goods;
➢
In order to save electrical energy, please try to reduce door-opening frequency and open time3.2
Temperature controller
3.2.1 Key function
The temperature controller is installed on the control panel of the medical refrigerator. When power is on, the
digital display screen can display actual temperature inside the refrigerator. The control panel is illustrated by
the following diagram:
❖
1.High temperature alarm indicator
❖
2.Door open indicator
❖
3..Defrost indicator