background image

 

 

 

  Notes:  If  other  parameter  need  to  be  changed,  please  contact  after-sale  or  local  distributor  of  us  for 

setting. 

  3.2.3 Error display 

  When  faults  of  the  refrigerator  temperature  display  sensor  including  short  circuit,  open  circuit, 

malfunction  and  incorrect 

connection  appear,  “E0”  is  displayed.  When  faults  of  the  refrigerator 

temperature  control  sensor  including  short  circuit,  open  circuit,  malfunction  and  incorrect  connection 

appear, “E1” is displayed. At the same time, the buzzer of the temperature controller will give out alarm 

and the display screen displays corresponding alarm parameters (E0

E1). 

  After the power is switched off, the control function of the controller fails and the back-up source is used 

for displaying. If the back-up source is charged for more than 10 hours, the power failure symbol 

Pd

the  current inside  temperature  and  the  alarm display  alternatively  for  one  time and  then display  again 

after 30S stop for cycle operation. After 8 hours, the nixie tube stops displaying. The power failure alarm 

light will 5S twinkle continuously until the electric quantity of the back-up source is used up. The inside 

temperature can be continuously displayed during the power failure. Please pay attention to this and take 

relevant measures to avoid damage. 

  Suggest:  If  you  judge  that  the  temperature  sensor  goes wrong,  you  should  cut  off  the  power  and  first 

check whether the connecting wire of the sensor is loose. Please contact professional after-sales service 

or distributor of us for processing. 

 

Tips: if internal code appears when user regulates the temperature controller, the user should 

wait  for  1  minute  and  then  operate  it  till  the  inside  temperature  is  displayed  to  avoid  fault  of  the 

medical refrigerator caused by the changing of the control parameter. 

  When  the  medical  refrigerator  stops  temporarily  due  to  power  failure  or  other  faults,  the  inside 

temperature  of  the  medical  refrigerator  will  rise  from  5

to  15

  within  1  hour.  If  power  cannot  be 

supplied in short time, the user should consider taking out the stored goods and putting them into other 

normal medical refrigerators for storing to avoid losses caused by the damage of the stored goods. 

  Before putting goods into the medical refrigerator to be stored, the user should confirm in advance that 

whether  inside  temperature  of  the  medical  refrigerator  is  in  accordance  with  the  set  temperature  for 

storing to avoid losses caused by goods damage due to the improper temperature. 

  Due  to  the  refrigeration  inertia,  the  medical  refrigerator  cannot  be  kept  at  constant  temperature.  And 

there is certain up and down difference between the inside temperature and the set temperature. The 

inside temperature varies with the changing of the environmental conditions and the temperature set . It 

belongs to normal phenomenon. 

Summary of Contents for MPC-5V100

Page 1: ...ON MANUAL Pharmacy Refrigerator MPC 5V100 VERSION 2020 10 MPC 5V100 The image shown here is indicative only If there is inconsistency between the image and the actual product the actual product shall...

Page 2: ...s goggles etc 2 Always adopt good health habits and operate strictly according to the product instruction 3 Everyone is obliged to be in charge of one own safety 4 Not Allowed direct contact with pati...

Page 3: ...perature The medical refrigerator realizes refrigeration through the heat dissipation of the back surface condenser In order to ensure normal operation and ventilation and heat dissipation the back an...

Page 4: ...3 2 1 Key function 7 3 2 2 Parameter settings 8 3 2 3 Error display 9 IV Routine Maintenance 10 4 1 Clean of incubator 10 4 2 Defrost 10 4 3 Care and maintenance 10 V Clearing of Fault and Maintenanc...

Page 5: ...oduct Diagram Structural diagram Due to the improvement and various models of products the actual product may be different from the diagram please in kind prevail Fan Shelf Glass door Door Lock Contro...

Page 6: ...mber Dismantle all package components including protection foam in the refrigerator body Please check accessories and data according to the packing list Please clean the product before use The product...

Page 7: ...ulator to be used cooperatively When the medical refrigerator is used the power supply is required to be equipped with a lower voltage air circuit breaker and a leakage protection device It is necessa...

Page 8: ...ed to put the goods to be stored in the refrigerator after the inside temperature falls to the operating temperature If there are too much goods to be stored it is advised to store separately for thre...

Page 9: ...mal working of the device such as bad temperature uniformity and abnormal high low temp alarm Notes the general high and low temperature alarm values are within the set temperature range of 2 10 3 2 2...

Page 10: ...emperature sensor goes wrong you should cut off the power and first check whether the connecting wire of the sensor is loose Please contact professional after sales service or distributor of us for pr...

Page 11: ...ged thereby causing faults During the rainy season vapors may be easily condensed on the surface of the glass door of the refrigerator In serious condition water may drip Please use dry cloth to wipe...

Page 12: ...cket is plugged or loosened If the power fuse is disconnected If the supply voltage is too low or too high Compressor breaks down If the temperature setting is right Temperature goes on decreasing aft...

Page 13: ...Main technical parameters 6 2 Electrical schematic diagram VII Packing list Model Temp Volume Power supply Refrigerant Dimension MPC 5V100 2 8 100L 220V 50Hz R600a 505 510 875 Model Operation instruc...

Reviews: