Medical refrigerator series products are applied to places needing storage at constant temperature, such as 

hospitals, epidemic prevention stations, blood banks, scientific research institutions, colleges and universities, 

biological pharmacies, genetic engineering, entrepreneurs, etc. The storing temperature in the body can be 

regulated by the temperature button on the control panel. The application is convenient and the performance 

is reliable. 

II. Preparation and Attention before Use 

  Transport:  the  refrigerator  should  be  uplifted  from  the  bottom  and  carried  and  put  down  lightly.  The 

inclined plane should be no larger than 45 degrees. 

  After installation, keep the device standing for 12 hours at least to return the compressor oil. 

  After installation and first use testing, please adjust the castors to fix the device. 

  Do not hold the door or the lining port as stressed member. 

  Dismantle all package components (including protection foam in the refrigerator body). 

  Please check accessories and data according to the packing list. 

  Please clean the product before use. 

  The product has self closing components, with door closed function, please pay attention when opening 

the door to pick up the goods. 

  Operating environment requirements: 

  For indoor use only; 

  The mounting surface must be fixed, horizontal and incombustible and be able to bear weight during the 

operation of the medical refrigerator; 

  To be placed away from direct glare of sunshine and heat and the environmental temperature should be 

not higher than 32

  Space  of  above  30cm  is  required  to  be  left  around  the  medical  refrigerator  for  ventilation  and  heat 

dissipation; 

  Not allowed to be placed in the environment under 0

℃;

 

  Not allowed to be placed at places with heavy moisture or easy-splashing water. 

  Ambient non corrosive, flammable, explosive gas, liquid or dust. 

  The surrounding environment must be kept well ventilated. 

  On flat ground, the medical refrigerator can be directly pushed to move. 

Summary of Contents for MPC-5V100

Page 1: ...ON MANUAL Pharmacy Refrigerator MPC 5V100 VERSION 2020 10 MPC 5V100 The image shown here is indicative only If there is inconsistency between the image and the actual product the actual product shall...

Page 2: ...s goggles etc 2 Always adopt good health habits and operate strictly according to the product instruction 3 Everyone is obliged to be in charge of one own safety 4 Not Allowed direct contact with pati...

Page 3: ...perature The medical refrigerator realizes refrigeration through the heat dissipation of the back surface condenser In order to ensure normal operation and ventilation and heat dissipation the back an...

Page 4: ...3 2 1 Key function 7 3 2 2 Parameter settings 8 3 2 3 Error display 9 IV Routine Maintenance 10 4 1 Clean of incubator 10 4 2 Defrost 10 4 3 Care and maintenance 10 V Clearing of Fault and Maintenanc...

Page 5: ...oduct Diagram Structural diagram Due to the improvement and various models of products the actual product may be different from the diagram please in kind prevail Fan Shelf Glass door Door Lock Contro...

Page 6: ...mber Dismantle all package components including protection foam in the refrigerator body Please check accessories and data according to the packing list Please clean the product before use The product...

Page 7: ...ulator to be used cooperatively When the medical refrigerator is used the power supply is required to be equipped with a lower voltage air circuit breaker and a leakage protection device It is necessa...

Page 8: ...ed to put the goods to be stored in the refrigerator after the inside temperature falls to the operating temperature If there are too much goods to be stored it is advised to store separately for thre...

Page 9: ...mal working of the device such as bad temperature uniformity and abnormal high low temp alarm Notes the general high and low temperature alarm values are within the set temperature range of 2 10 3 2 2...

Page 10: ...emperature sensor goes wrong you should cut off the power and first check whether the connecting wire of the sensor is loose Please contact professional after sales service or distributor of us for pr...

Page 11: ...ged thereby causing faults During the rainy season vapors may be easily condensed on the surface of the glass door of the refrigerator In serious condition water may drip Please use dry cloth to wipe...

Page 12: ...cket is plugged or loosened If the power fuse is disconnected If the supply voltage is too low or too high Compressor breaks down If the temperature setting is right Temperature goes on decreasing aft...

Page 13: ...Main technical parameters 6 2 Electrical schematic diagram VII Packing list Model Temp Volume Power supply Refrigerant Dimension MPC 5V100 2 8 100L 220V 50Hz R600a 505 510 875 Model Operation instruc...

Reviews: