4.2.3 MOUNT FITTING
Fig. 18.0
4. Lift Phantom arm into place, and engage the 24V plugs as shown in Fig. 18.0.
5. Insert the arm bearing into the wall unit (taking care to align the threaded holes on the arm side
bearing sheath with the corresponding holes on the bearing sheath on the wall unit, as shown in Fig.
18.0.
6. Insert and tighten the two grub screws to complete the fixing of the arm to the wall unit as shown
in Fig.18.0.
7. Replace the wall unit cover as shown in Fig. 18.0.
8. Fix in place with the countersunk screws provided, as shown in Fig. 18.0.
MAINTENANCE ONLY
9. Unscrew the end cap for access to the spring balance adjustment, as shown in Fig. 18.0.
10. Adjust the spring balance: using a 4mm allen key, adjust the screw inside the knuckle. Clock-
wise/Tightening the screw will cause heavier engagement of the spring - this will improve ‘head
droop’. Anticlockwise/loosening the spring balance will fix a ‘rising head’ issue, as shown in Fig 18.0.
11. Friction adjustment: Clockwise/tightening this screw will apply more friction to the arm mecha-
nisms, as shown in Fig. 18.0. The arm will require more force to move, however will not continue to
move with moment. This should be set to the staff preference.