background image

 

PN 350-0453000-02 

 

P a g e

 | 

4

 

 

 

 

 

 

Aligning the FlexIR Accessory  

 
This accessory has been pre-aligned on the same make as your spectrometer. In most cases, little to no 
further alignment is necessary. To assess the need for additional alignment, check the energy when in 
the monitor mode of the spectrometer software. When using the specular or diffuse reflection probe, 
slide the reference cap over the probe tip before monitoring the energy. No reference cap required for 
ATR probes. 
 
The alignment procedure requires that the IR beam pass through the hollow waveguide(s) and return 
the beam through the hollow waveguide to reach the detector. Prior to adjusting the mirrors as 
explained below a closed loop must be established. In other words, the alignment process for a diffuse 

reflectance or a specular reflectance probe requires that the probe tip be positioned gently touch

ing the 

diffuse reference disk or a specular mirror embedded in the reference cap included with the probe, 
respectively. For an ATR probe, no reference or mirror is required. Simply be sure the ATR crystal surface 
is clean.  
 
Open the FTIR bench setup software display to monitor the interferogram signal strength. Please refer 
to your FTIR software manual for details of this operation. Reduce the size of the FTIR aperture to 
achieve acceptable interferogram signal strength. If your FTIR does not have an aperture, use an IR 
beam-limiting screen placed at the right or left side of the FTIR sample compartment to reduce the 
energy to an acceptable level.  
 
Optimize the signal by adjusting the input mirror (Figure 3), which directs the IR beam into the hollow 
waveguide(s), located below the installed probes. Rotate slowly the screws, positioned on the diagonal, 
one at a time while maximizing the signal by monitoring the energy displayed through the FTIR software.  

Figure 2: Assembled FlexIR Accessory

 

 

Summary of Contents for FlexIR Hollow Waveguide

Page 1: ...ocument contains confidential or proprietary information of PIKE Technologies Inc Neither this document nor the information herein is to be reproduced distributed used or disclosed either in whole or...

Page 2: ...ilization 3 Installing the HWG Probe 3 Aligning the FlexIR Accessory 4 Optimizing MCT Pre Amp Gain 6 Removing the FlexIR Accessory 6 Storage 6 ATR Probe 7 Zinc Selenide 7 Germanium 7 Diamond ZnSe Comp...

Page 3: ...oviding exceptional IR throughput The new hollow waveguides are very durable and free from the typical fracture problems encountered with polycrystalline and glass core fibers The highly reflective HW...

Page 4: ...hollow waveguide probes have a limited bend radius of 150 mm A smaller radius bend will break the probe Packing List FlexIR Manual FlexIR Hollow Waveguide Base HWG Probe or Probes PN 045 3000 PN 045...

Page 5: ...nt of liquid nitrogen into the funnel and wait for it to drain into the Dewar of the detector Add additional small amounts of liquid nitrogen until the detector is filled Allow the detector to stabili...

Page 6: ...probe tip be positioned gently touching the diffuse reference disk or a specular mirror embedded in the reference cap included with the probe respectively For an ATR probe no reference or mirror is r...

Page 7: ...ment of the input and output mirror two to three more times if increases in performance are observed to fully optimize the accessory The positioning of the MCT detector has been optimized for the spec...

Page 8: ...in should be adjusted to allow 50 of total energy counts or voltage registering on the detector To adjust the gain remove the gain cap from the front of the base optics Access the gain screw using a s...

Page 9: ...dditionally complexing agents such as ammonia and EDTA will erode its surface because of the formation of complexes The maximum temperature range for the ZnSe probe is ambient to 95 C Germanium German...

Page 10: ...n diffuse surfaces In collecting a background spectrum insert probe into provided reference cap If probe fits tightly or cannot bottom out on reference loosen set screw on side and insert probe cap Ti...

Page 11: ...pectrum of a coating on a smooth reflective surface coating on the outside of a soda can conducive to specular reflectance measurements The second sample featured a painted surface with diffuse reflec...

Page 12: ...ld be used when handling and using ATR crystals since some of the materials can be hazardous Speci cally zinc selenide is a heavy metal material and should be handled with this in mind If the crystal...

Page 13: ...wheeled bin is a clear reminder that the product must NOT be disposed with household waste It is the responsibility of the buyer to discard the product in accordance with Federal regional and local e...

Reviews: