been turned off and the default value will be restored. You can set the
number of samples on each trigger through front panel or the remote
interface.
Front Panel Operation
1.
Press SHIFT+CONFIG button.
2.
Set a defined parameter via the procedure TRIG SYS > TRIG CNT >
ENTER.
3.
Set the desired number by pressing
△
and
▽
buttons to increase or
decrease the number, and using
◁
and
▷
buttons to move
through the digits. Press ENTER to set the number.
Remote Interface Operation
Use the following command to set the number of samples from your PC
terminal.
SAMPle:COUNt <value>
B. Number of triggers
Although the meter normally takes one trigger before returning to the
“idle” state, you can manually specify the number of triggers it accepts
before the “idle” state. However, this can only be done through the
remote interface. The following command shows how to set multiple
triggers before returning to idle state.
TRIGger:COUNt <value>
C. Reading hold(Front panel operation only)
The reading hold feature is used to hold a stable reading on the display.
When a reading is stable, enabling the reading hold will hold the stable
reading and trigger a beeping sound. Reading hold feature depends on
an adjustable sensitivity band to determine if a reading is stable. The
multimeter considers a reading stable when three consecutive readings
are within the sensitivity band. Chapter 4.4.6 will explain how to adjust
the sensitivity band. This feature is only available from the front panel
operation.