138
138
138
138
decimal value which corresponds to the binary-weighted sum of all bits
set in the register.
STATus:PRESet
Clear all bits in the Questionable Data enable register.
*CLS
Clear the Status Byte summary register and all event registers.
*ESE <enable value>
Enable bits in the Standard Event enable register. The selected bits are then
reported to the Status Byte.
*ESE?
Query the Standard Event enable register. The multimeter returns a
decimal value which corresponds to the binary-weighted sum of all bits
set in the register.
*ESR?
Query the Standard event register. The multimeter returns a decimal
value which corresponds to the binary-weighted sum of all bits set in the
register.
*OPC
Sets the “operation complete” bit (bit 0) in the Standard Event register after the
command is executed.
*OPC?
Returns “1” to the output buffer after the command is executed.
*PSC {0|1}
Power-on status clear. Clear the Status Byte and Standard Event
register enable masks when power is turned on (*PSC 1). When *PSC 0
is in effect, the Status Byte and Standard Event register enable masks
are not cleared when power is turned on. [Stored in non-volatile memory]
*PSC?
Query the power-on status clear setting. Returns “0” (*PSC 0) or “1” (*PSC 1).