TRIGger:DELay {<seconds>|MINimum|MAXimum}
or
TRIGger:DELay:AUTO {OFF|ON}
4.3 Math Operations
This section will introduce the mathematical operations in M3510A/11A.
There are eight math operations: PERCENT, AVERAGE, NULL, LIMITS
(HIGH LIMIT/LOW LIMIT), MX+B, dB and dBm testing. They either store
data for later use or perform mathematical operations on the readings.
Note that these math operations are available to all measurement
functions except diode and continuity testing.
※
Note: The “MATH” anunnciator on the display indicates the state of a
mathematical feature, excluding Ratio, and will be lit when enabling
whichever math feature.
4.3.1 Percent
Definition
This mathematical function calculates the ratio of a measurement
reading to a specified target value as the form of percentage. The
calculation formula is shown below:
The specified target value is stored in a volatile memory and will be
cleared after the meter has been turned off or a remote interface reset.
How to use Percent function
There are two ways to make a percent measurement: Through the front
panel operation or through the remote interface operation.