can set the address to any value from 0 and 31, and the setting will be
stored in the non-volatile memory. The address is set to “22” when the
multimeter is shipped from the factory.
※
Note: The remote interface can only be set through the front panel
operations.
How to select a remote interface
1.
Press SHIFT + CONFIG buttons. Then use
◁
and
▷
buttons to
locate “INTERFACE”. Press ENTER to select it.
2.
Use
◁
and
▷
buttons to toggle between “USB” and “GPIB”. Press
ENTER on your choice of interface.
How to set address when using GPIB
1.
Press SHIFT + CONFIG buttons, and then use
◁
and
▷
buttons to
locate “INTERFACE”. Press ENTER to select.
2.
Use
◁
and
▷
buttons to locate “GPIB”. Press ENTER to select it.
3.
Use
◁
and
▷
buttons to move through the digits and
△
and
▽
buttons to adjust the numbers to a desired GPIB address. Press
ENTER to set the address.
4.2 Trigger Operations
In this section we will discuss the triggering system in M3510A/11A.
M3510A/11A provides a variety of trigger operations. You can select a
trigger mode, a trigger source and different trigger settings for a specific
measurement. The defined selection is stored in a volatile memory and
the default settings will be restored after power-off.
Figure 4-1
shows
the trigger operation in M3510A/11A.