About the Standard Event Register
The standard event register reports the instrument event types below,
such as power-on detected, command syntax errors, command
execution errors, self-test (calibration errors), query errors or the
moment of executing an *OPC command. Then through the enable
register, all conditions will be reported in the standard event summary
bit. At the moment, users have to write a decimal value by using the
*ESE command to set the enable register mask.
※
Note 1: The errors in the DMM’s error queue except reading the error
queue by using SYSTem:ERRor? will be recorded by an error condition
(including bits 2, 3, 4 or 5 of the standard event register).
※
Note 2: In both the standard event register bit 3 and the questionable
data event register bits 0, 1 or 9, a reading overload condition can be
reported all the time. But in the DMM’s error queue, no error message
will be recorded.
Bit Definitions for Standard Event Register
Bit
Decimal
Value
Definition
0.
Operation Complete
1
All commands prior to and including an *OPC command have
been executed
1.
Not Used
2
Set to 0.
2.
Query Error
4
The DMM output buffer was empty as the DMM tried to read from
it. Or a new command line was received before a previous query
has been read. Or, both the input and output buffers are full.
3.
Device Error
8
Error occurrence from a self-test, calibration or reading overlaod.
4.
Execution Error
16
Error occurrence from an execution.
5.
Command Error
32
Error occurrence from a command’s syntax.
6.
Not Used
64
Set to 0.
7.
Power On
128
Since the last time the event register was read or cleared, power