Command
Parameter
Description
gcur
-
returns the actual setpoint current in [A].
gcurmin
-
returns the minimum setpoint current.
gcurmax
-
returns the maximum setpoint current.
scur
<current>
sets the internal setpoint current to the given value. This
value must be within the minimum/maximum borders
(see above). One position after decimal point is used
(e.g. 80.5). The return value is the new setpoint.
gwidth
-
returns the actual pulse width in [us].
gwidthmin
-
returns the minimum possible pulse width in [us].
gwidthmax
-
returns the maximum possible pulse width in [us].
swidth
<width>
sets the pulse width for the pulse.
Please note that any change in this register affects the
maximum possible repetition rate. The return value is the
new pulse width.
greprate
-
returns the actual repetition rate in [Hz]
grepratemin
returns the minimal possible repetition rate in [Hz]
grepratemax
returns the maximal possible repetition rate in [Hz]
sreprate
reprate in [Hz]
sets the pulse repetition rate to the given value. The
return value is the new repetition rate.
gvcap
-
returns the actual pre charge voltage of the internal
capacitor bank in [V]
gvcapmin
-
returns the minimum pre charge voltage of the internal
capacitor bank in [V]
gvcapmax
-
returns the maximum pre charge voltage of the internal
capacitor bank in [V]
svcap
voltage in [V]
sets the pre charge voltage of the internal capacitor bank
to the given value in [V]
gffwd
-
returns the actual voltage of the feed forward part of the
current regulator in [V]. See section “Current Regulator”
for more information.
sffwd
<voltage>
sets the feed forward voltage to the given value. Two
positions after decimal point are used (e.g. 3.45). It must
be within the borders of gffwdmin / gffwdmax.
gffwdmin
-
returns the minimum possible value of the feed forward
voltage in [V]
gffwdmax
-
returns the maximum possible value of the feed forward
voltage in [V]