Description of the ERROR Registers
The device has two 32-bit wide error registers. The following lists contain a description of the
individual bits of each register. Unless otherwise noted, a “1” indicates an error or warning
condition.
Bit
Name
Read/
Write
Meaning
0
CRC_DEVDRV_FAIL
ro
A CRC error was detected in the internal
PLB-21 driver. Please contact your
distributor.
1
CRC_DEFAULT_FAIL
ro
A CRC error was detected in the default
values. A re-save of the values should
correct this.
2
CRC_CONFIG_FAIL
ro
A CRC error was detected in the internal
configuration values. Please contact your
distributor.
3
reserved
ro
reserved
4
CRC_FFWDCAL_FAIL
ro
Indicates that the feed forward calibration
of channel 1 is faulty.
5
CRC_ISOLLCAL_FAIL
ro
Indicates that the current setpoint
calibration of channel 0 is faulty.
6
TEMP_OVERSTEPPED
ro
The internal temperature was beyond safe
operating limits.
7
TEMP_WARNING
ro
The internal temperature is 5 °C before
shutdown.
8
TEMP_HYSTERESE
ro
Device is cooling down. Temperature
needs to drop below (maximum – 10 °C)
9
VCC_FAIL
ro
Internal supply voltage error.
10
FAIL_DEFAULTS
ro
The loading of the default failed. Normally
this is because of a pending CRC error.
11
I2C_EEPROM_FAIL
ro
Internal EEProm error. Please contact your
distributor.
12
I2C_DAC_FAIL
ro
Internal DAC error.
13
I2C_RD_FAIL
ro
Indicates an I²C read error.
14
I2C_WR_FAIL
ro
Indicates an I²C write error.
15
ENABLE_POWERON
ro
Indicates that the ENABLE signal was
given during start-up.
16
TEMP_SENSOR_FAIL
ro
A “1” indicates an error.
17-31 reserved
ro
Reserved