Rev. 01/2020
35
Default values
These commands give access to the driver’s default value set. Each operating parameter can
be saved to this default set and recalled later.
Feed Forward (FFwd)
These commands give access to the current regulators feed forward part. These commands are
only used, if the current regulator is set to mode 0 and should be used with caution. If the
regulator is in any other mode, the answer frame will contain the command 0xFF14 indicating
that the command is unavailable.
Command Name
Command frame
Answer frame
Command
Data
Command Data
LOADDEFAULTS
0x0800
0
0x0800
0
SAVEDEFAULTS
0x0801
0
0x0800
0
Command Name
Command frame
Answer frame
Command
Data
Command Data
GETFFWD
0x1000
0
0x9000
Uint32 in 0.01 V
SETFFWD
0x1001
Uint32 in 0.01 V
0x9000
Uint32 in 0.01 V
GETFFWDMIN
0x1002
0
0x9000
Uint32 in 0.01 V
GETFFWDMAX
0x1003
0
0x9000
Uint32 in 0.01 V