Rev. 01/2021
22
Defaults
The driver can load a default setting each time it powers up or the user commands it to do so.
This is done within this submenu.
Def. pwron
When enabled, the driver loads the saved settings each time it powers up.
Load
When activated via turning the Jog-dial or the ENTER key all internal registers are changed
to the previously saved values. The output stage has to be re- enabled via the L_ON bit or the
FIRE Key afterwards.
Save
When activated via turning the Jog-dial or the ENTER key all internal registers are stored
into an internal EEprom for later usage.
Temperature
The LDP-C /-CW is equipped with several temperature sensors. In this submenu, the actual
and shutdown temperatures can be read and modified. All values are in °C.
Dev. Off
This shows the user defined shutdown temperature. If the LDP-C /-CW reaches this
temperature during operation, the output will be disabled and an error message is displayed.
It can be modified within
40 .. 80 °C.
Temp 1…3
This shows the actual temperatures measured by the sensors of the LDP-C /-CW.
Measurement
This submenu holds some runtime measurement information.
Uin
This shows the measured value of the supply voltage.
Udiode
This shows the measured value of the compliance voltage of the connected load. The output
must be enabled for a correct result. Please note that this is not a real time measurement. If
the setpoint is modulated, the display may show incorrect values. In that case use the signal
of the analog current monitor at pin 9 on the BOB connector with appropriate measurement
equipment.
Idiode
This shows the measured value of the current flow through the connected load.