Suitable Electronics
Model /
Mechanics
Model / Electronics
Capacitive sensor,
sub-D connection:
P-725.xCD
P-725.xCA
P-725.CDD
E-621 controller module
E-625 servo controller, bench-top
E-665 display servo controller, with digital interface, bench-top
E-754 single-channel digital controller
E-709 single-channel digital controller
Capacitive sensor,
LEMO connection:
P-725.xCL
E-610 servo controller / amplifier
E-500 modular piezo controller system with E-505 high-performance
amplifier module and E-509 controller
Strain gauge
sensor, LEMO
connection:
P-725.SDD
E-610 servo controller / amplifier
E-625 servo controller, bench-top
E-665 display servo controller, with digital interface, bench-top
E-709 single-channel digital controller
No sensor,
LEMO connection:
P-725.x0L
E-500 modular piezo controller system with E-505 high-performance
amplifier module
E-610.00 amplifier
Installation
Connecting the P-725 to the Protective Earth Conductor
INFORMATION
Observe the applicable standards for mounting the protective earth conductor.
The P-725 is equipped with an M4 hole for fastening the protective earth conductor. This hole is
marked with the protective earth conductor symbol
(see " Dimensions", p. 19).
Requirements
You have read and understood the safety precautions (p. 3).
The P-725 is
not
connected to the electronics.
Tools and accessories
Suitable protective earth conductor: Cross-
sectional area of the cable ≥0.75 mm
2
Supplied M4 protective earth screw set (p. 7) for connecting the protective earth conductor
Suitable screwdriver