Page 12 / 25
2018-11-13
P725T0010, valid for P-725
Physik Instrumente (PI ) GmbH & Co. KG, Auf der Roemerstr. 1, 76228 Karlsruhe/Germany
Phone +49 721 4846-0, Fax +49 721 4846-1019, Email [email protected],
www.pi.ws
Installing the Objective
Correct:
Figure 4: Center of load close to aperture
Incorrect:
Mounting the load incorrectly causes high strain on the flexure guides in the Z drive, high torques and
the danger of oscillations.
Figure 5: Center of load far below the aperture
Requirements
You have read and understood the safety precautions (p. 3).
You have affixed the P-725 to the microscope (p. 10).
Tools and accessories
P721T0002 technical note for PIFOC QuickLock thread option (p. 7)
QuickLock thread adapter (p. 8) (not included)
Objective to be installed (not included)
Suitable tools
Installing the objective
Screw the objective into the P-725. See the P721T0002 technical note (p. 7) for details.
Pay attention to a maximum torque of 0.6 Nm.