INFORMATION 

 

Systems are calibrated at the factory to achieve optimum performance. Replacing 
the system components will cause a loss in performance when Z drives without 
ID chip are used, or whose ID chip does not contain any calibration data, or when 
Z drives with LEMO connectors or analog controllers are used. 

Only P-725.xxD and P-725.xCA Z drives are equipped with an ID chip. 

  Note the assignment of the Z drive axis to the controller channel, which is given 

by the calibration label of the piezo servo controller. 

  For systems without ID chip: 

If the piezo servo controller or the Z drive has to be replaced, recalibrate the axis 
displacement (see controller manual) or contact our customer service 
department (p. 15). 

 
 

Starting Up and Operating the P-725 

Requirements 

  You have read and understood the safety precautions (p. 3). 

  You have read and understood the general notes on startup and operation (p. 13). 

Starting up and operating the P-725 

  Follow the instructions in the user manual of the piezo controller used for startup and 

operation of the P-725. 

 
 

Discharging the P-725 

The P-725 must be discharged before demounting. Demounting is necessary e.g., before cleaning or 
transporting the P-725 as well as for modifications 

Discharging a P-725 that is connected to the controller 

In closed-loop operation: 

1.  Switch off the servo mode on the controller. 

2.  Set the piezo voltage to 0 V on the controller. 

In open-loop operation: 

  Set the piezo voltage to 0 V on the controller. 

Discharging a P-725 that is not connected to the controller 

  Connect the P-725 to the switched-off PI controller for 10 seconds. 

 
 

Summary of Contents for PIFOC P-725 series

Page 1: ...lid for P 725 Physik Instrumente PI GmbH Co KG Auf der Roemerstr 1 76228 Karlsruhe Germany Phone 49 721 4846 0 Fax 49 721 4846 1019 Email info pi ws www pi ws P 725 PIFOC Long Travel Objective Scanner...

Page 2: ...ing the P 725 to the Protective Earth Conductor 9 Affixing the P 725 to the Microscope 10 Installing the Objective 12 Startup and Operation 13 General Notes on Startup and Operation 13 Starting Up and...

Page 3: ...r understandability the colors proportions and degree of detail in illustrations can deviate from the actual circumstances Photographic illustrations may also differ and must not be seen as guaranteed...

Page 4: ...ock max 120 V DC and minor injuries Do not touch the contacts in the connector Secure the connector of the piezo actuator with screws against being pulled out of the controller CAUTION Risk of electri...

Page 5: ...adapters P 725 1CL PIFOC piezo nanofocusing Z drive for long scanning ranges 100 m capacitive sensors LEMO connectors for QuickLock thread adapters P 725 10L PIFOC piezo nanofocusing Z drive for long...

Page 6: ...400 m capacitive sensors LEMO connectors for QuickLock thread adapters P 725 40L PIFOC piezo nanofocusing Z drive for long scanning ranges 460 m open loop LEMO connectors for QuickLock thread adapters...

Page 7: ...nsecutive numbers PIFOC Brand name Country of origin Germany Country of origin Warning sign Observe manual Old equipment disposal p 25 WWW PI WS Manufacturer s address website CE conformity mark Manuf...

Page 8: ...hread adapter M32 x 0 75 mm 21 mm P 721 06Q QuickLock thread adapter M26 x 1 36 21 mm P 721 08Q QuickLock thread adapter M19 x 0 75 mm 14 mm P 721 11Q QuickLock thread adapter M25 x 0 75 mm 21 mm P 72...

Page 9: ...rvo controller amplifier E 625 servo controller bench top E 665 display servo controller with digital interface bench top E 709 single channel digital controller No sensor LEMO connection P 725 x0L E...

Page 10: ...onnection of the P 725 as shown in the profile view 3 Tighten the M4 screw with a torque of 1 2 Nm to 1 5 Nm 4 Make sure that the contact resistance at all connection points relevant for mounting the...

Page 11: ...ions p 3 Tools and accessories P721T0002 technical note for PIFOC QuickLock thread option p 7 QuickLock thread adapter p 8 not included Note that you only need the objective insert of the QuickLock th...

Page 12: ...e Z drive high torques and the danger of oscillations Figure 5 Center of load far below the aperture Requirements You have read and understood the safety precautions p 3 You have affixed the P 725 to...

Page 13: ...y of the air Avoid operating the P 725 in environments that can increase the electric conductivity Only operate the P 725 within the permissible ambient conditions and classifications p 25 NOTICE Dest...

Page 14: ...be replaced recalibrate the axis displacement see controller manual or contact our customer service department p 15 Starting Up and Operating the P 725 Requirements You have read and understood the sa...

Page 15: ...an the surface of the P 725 with a cloth that is lightly dampened with a mild cleanser or disinfectant e g ethanol or isopropyl alcohol Do not do any ultrasonic cleaning Customer Service For inquiries...

Page 16: ...pacitive Capacitive Open loop travel 20 to 120V 150 330 460 as P 725 xCL m min 20 0 Closed loop travel 100 250 400 m calibrated Open loop resolution 0 3 0 4 0 5 as P 725 xCL nm typ Closed loop resolut...

Page 17: ...pe PICMA P 885 PICMA P 885 PICMA P 885 as P 725 xCL Electrical capacitance 4 2 6 2 6 2 as P 725 xCL F 20 Dynamic operating current coefficient 5 2 3 1 1 9 as P 725 xCL A Hz m 20 Miscellaneous Operatin...

Page 18: ...lity 1 5 5 nm typ Runout X Y 2 2 rad typ Crosstalk in X Y 150 150 nm typ Mechanical properties Stiffness in motion direction 1 5 1 5 N m 20 Resonant frequency no load 1180 1180 Hz 20 Resonant frequenc...

Page 19: ...76228 Karlsruhe Germany Phone 49 721 4846 0 Fax 49 721 4846 1019 Email info pi ws www pi ws Dimensions Dimensions in mm Note that the decimal places are separated by a comma in the drawings P 725 xCD...

Page 20: ...valid for P 725 Physik Instrumente PI GmbH Co KG Auf der Roemerstr 1 76228 Karlsruhe Germany Phone 49 721 4846 0 Fax 49 721 4846 1019 Email info pi ws www pi ws P 725 xCA Figure 7 P 725 xCA P 721 xxA...

Page 21: ...2018 11 13 P725T0010 valid for P 725 Physik Instrumente PI GmbH Co KG Auf der Roemerstr 1 76228 Karlsruhe Germany Phone 49 721 4846 0 Fax 49 721 4846 1019 Email info pi ws www pi ws P 725 xDD Figure 8...

Page 22: ...L bottom 1 LEMO coaxial connector for capacitive sensor Probe 2 LEMO coaxial connector for capacitive sensor Target 3 LEMO coaxial connector for piezo voltage LEMO coaxial connectors Figure 10 LEMO co...

Page 23: ...ws www pi ws P 725 SDD Figure 11 LEMO connectors of P 725 SDD 1 4 pin LEMO connector for strain gauge sensor 2 LEMO coaxial connector for piezo voltage Sensor connector Figure 12 Sensor connector fron...

Page 24: ...sensor A2 outer conductor GND Shield 1 Bidirectional Data line for ID chip 2 GND Shield of Target Ground of ID chip when switched on 3 Input Piezo voltage 4 N C Not connected 5 Input Target sensor si...

Page 25: ...temperature 25 C to 85 C Overvoltage category II Protection class I Degree of pollution 1 Degree of protection according to IEC 60529 IP20 Old Equipment Disposal In accordance with EU law electrical...

Reviews: