3 Product Description 

Version: 1.2.0 

MP45E 

M-605 High-Precision Linear Stage 

3.4

 

Technical Features 

3.4.1

 

Linear Scale Position Encoder 

M-605 stages use linear scale position encoders with 0.1 µm resolution for position detection. 

Maximum velocities of 50 mm/s can be achieved. 
The linear encoder is mounted in the center of the stage, close to the ballscrew drive to prevent 

possible cosine error. The linear encoder measures the actual position of the moving carriage 

directly, thus eliminating drivetrain errors such as non-linearity, backlash or elastic 

deformations. 
The encoder is based on an optical grating and a non-contacting read head with integrated 

signal conditioner and interpolator. 

3.4.2

 

PWM Amplifiers 

For maximum dynamic performance, the DC servo-motors are driven by high-efficiency PWM 

power amplifiers integrated into the stages. An external line-power power supply (order 

number: M-500.PS) is provided to supply the built-in amplifiers directly. This architecture allows 

high torque and high velocities while loading the motor controller with control signals only. The 

actual power is provided by the external supply. 

3.4.3

 

Travel Limit Sensors 

M-605 stages are protected against running into the hard stop by double-level magnetic limit 

sensors (Hall-effect sensors with TTL drivers) at each end of travel. 
The 

inner limit sensors

 (N1 at the negative-travel end and P1 at the positive-travel end) work 

with the controller's limit sense input lines. The TTL output signal is active high. 
The 

outer limit sensors

 (N2 at the negative and P2 at the positive end) work locally, opening a 

relay that cuts the motor current. The second-level limit switches provide for fail-safe operation 

in case the controller fails to stop the motor when the inner limit sensor is reached. 
If the outer limit switch is reached, the stage can not be operated by the controller until the 

platform is moved manually. Disconnect the motor cable or set the motor into MOTOR OFF 

state before moving the platform back into the current-allowed area via the rotary knob. 

3.4.4

 

Position Reference Signal Sensors 

Position Reference Sensors are located approximately in the middle of the operating range and 

can be used to reference the absolute position of the stage. Always approach the reference 

sensor from the same side to reach the same position. 
The reference sensor in M-605 stages provides a static signal level which depends on the 

platform position. If the platform is on the “positive side” the reference signal is +5 V, while if 

the platform is on the “negative side,” the signal level is 0 V. Most PI motor controllers and 

Windows libraries offer the option of starting a search run for the reference point using the 

current reference sensor signal to determine the appropriate direction. 
The "AutoFindReference" function allows for starting a search run for the reference signal. 

Independent at which side relative to the reference position the stage is located, this function 

always starts the stage towards the reference position. 

Summary of Contents for M-605 Series

Page 1: ...MP45E M 605 High Precision Linear Stage User Manual Version 1 2 0 Date 2019 07 02 This document describes the following products M 605 1DD Compact precision linear stage 25 mm travel range 0 1 m linea...

Page 2: ...9 Physik Instrumente PI GmbH Co KG Karlsruhe Germany The text photographs and drawings in this manual are protected by copyright With regard thereto Physik Instrumente PI GmbH Co KG reserves all right...

Page 3: ...livery 5 3 3 Accessories 5 3 4 Technical Features 6 3 4 1 Linear Scale Position Encoder 6 3 4 2 PWM Amplifiers 6 3 4 3 Travel Limit Sensors 6 3 4 4 Position Reference Signal Sensors 6 4 Unpacking 7 5...

Page 4: ...able 21 10 1 2 Maximum Ratings 22 10 1 3 Ambient Conditions and Classifications 22 10 1 4 Limit Switch Specifications 23 10 2 Dimensions 24 10 2 1 M 605 1DD 24 10 2 2 M 605 2DD 25 10 2 3 M 605 AV1 26...

Page 5: ...nual CAUTION Dangerous situation If not avoided the dangerous situation will result in minor injury Actions to take to avoid the situation NOTICE Dangerous situation If not avoided the dangerous situa...

Page 6: ...the data storage device for the product to get the access data Downloading manuals 1 Open the website www pi ws 2 If the product was shipped with a data storage device Log into the website a Click Log...

Page 7: ...amage to the M 605 Only use the M 605 for its intended purpose and only if it is in perfect condition Read the user manual Immediately eliminate any faults and malfunctions that are likely to affect s...

Page 8: ......

Page 9: ...ecision linear stage 50 mm travel range 4295 Mounting kit consisting of 4 metric screws M4x35 and a hex wrench C 815 38 Motor cable D Sub 15 3 m C 501 24050H Wide range input power supply 24 V DC 50 W...

Page 10: ...end work with the controller s limit sense input lines The TTL output signal is active high The outer limit sensors N2 at the negative and P2 at the positive end work locally opening a relay that cut...

Page 11: ...ith care 2 Compare the contents with the items listed in the contract and the packing list 3 Inspect the contents for signs of damage If there is any sign of damage or missing parts contact PI immedia...

Page 12: ......

Page 13: ...s 5 2 Mounting the M 605 M 605 stages can be mounted in any orientation horizontally or vertically To achieve the specified guiding accuracy the stages must be mounted on an even surface to avoid tors...

Page 14: ...You have prepared the load so that it can be affixed to the mounting holes on the upper platform p 24 Tools and accessories At least 2 metric screws M4x35 Hex wrench in the scope of delivery p 5 Affix...

Page 15: ...ents The power cord is not connected to the power socket Tools and accessories Supplied components in the scope of delivery p 5 Wide range input power supply Adapter for the power supply connection ba...

Page 16: ......

Page 17: ...be caught by moving parts Stop the motion immediately if a controller malfunction occurs If possible adapt the travel range limits of your mechanical system in the software you use for commanding the...

Page 18: ...ler and the required PC software have been installed Starting and Operating the M 605 1 Start the controller see user manual for the controller 2 Configure the controller for the M 605 during startup...

Page 19: ...he stages are operated in extremely dusty environments we recommend cleaning and greasing the ballscrew and linear bearings from time to time The time interval depends on the degree of contamination a...

Page 20: ......

Page 21: ...p 11 Fault Reduced positioning accuracy Possible causes Remedial measures Warped base body Mount the M 605 onto an even surface p 9 Fault Uncontrolled oscillation Possible causes Remedial measures La...

Page 22: ......

Page 23: ...serial numbers of all products in the system Firmware version of the controller if applicable Version of the driver or the software if applicable Operating system on the PC if applicable If possible...

Page 24: ......

Page 25: ...lution 0 1 0 1 m Design resolution 0 1 0 1 m Minimum incremental motion 0 3 0 3 m Unidirectional repeatability 0 1 0 1 m Bidirectional repeatability 0 2 0 2 m Accuracy 1 1 m Pitch 30 30 rad Yaw 30 30...

Page 26: ...es 10 1 2 Maximum Ratings The M 605 positioners are designed for the following operating data Maximum operating voltage Maximum operating frequency Maximum power consumption 24 V 6 W 10 1 3 Ambient Co...

Page 27: ...h Specifications Type Magnetic Hall effect sensors can be used with bipolar CMOS circuits Supply voltage 5 24 V GND supplied by the motor controller through the motor connector Signal output TTL level...

Page 28: ...10 Technical Data 24 Version 1 2 0 MP45E M 605 High Precision Linear Stage 10 2 Dimensions Dimensions in mm Note that the decimal places are separated by a comma in the drawings 10 2 1 M 605 1DD...

Page 29: ...10 Technical Data M 605 High Precision Linear Stage MP45E Version 1 2 0 25 10 2 2 M 605 2DD...

Page 30: ...10 Technical Data 26 Version 1 2 0 MP45E M 605 High Precision Linear Stage 10 2 3 M 605 AV1...

Page 31: ...Signal Direction 1 Control for optional motor brake 0 to 24 V Input 2 Not connected 3 PWM MAGN Input 4 5 V Input 5 PLIMIT positive limit switch Output 6 GND Limit 7 Encoder A Output 8 Encoder B Output...

Page 32: ......

Page 33: ...g to international national and local rules and regulations In order to fulfil the responsibility as the product manufacturer PI miCos GmbH undertakes environmentally correct disposal of all old PI mi...

Page 34: ......

Page 35: ...M 605 in accordance with the following European directives EMC Directive Low Voltage Directive Safety of Machinery The standards applied for certifying the conformity are listed below EMC EN 61326 1...

Reviews: