4. Make sure that the contact resistance is <0.1
Ω
at 25 A at all connection points relevant for
attaching the protective earth conductor.
6.3
Mounting the Load onto the L-417.099232B
Overview
Figure 3: Mount the load on the L-417.099232B
1. Screws
2. Load
3. Motion platform of the L-417.099232B
Tools and Accessories
■
At least 3 screws with suitable
dimensions (p. 30)
■
Suitable tool for tightening the screws
■
Optional: 2 suitably dimensioned dowel pins as locating pins for aligning the load on the
L-417.099232B
Requirements
✓
You have read and understood the
general safety instructions (p. 7)
.
✓
You have mounted the L-417.099232B
on the surface (p. 13)
properly.
✓
The L-417.099232B is not connected to the electronics.
✓
You have prepared the load so that it can be fixed to the mounting holes on the motion
platform:
■
The distance between the center of gravity of the load and the center of the motion
platform is as small as possible in all directions.
■
At least three points are provided for fixing the load on the motion platform.
6 INSTALLATION
L417M0004EN
‒
11/21/2018
16
M O T I O N | P O S I T I O N I N G