C.4 Look-Up Table (LUT)
C.4
Look-Up Table (LUT)
C.4.1
Overview
The LUT is described in detail in Section 11.4.
To manually set custom LUT values in the GUI is practically not feasable as up to
4096 values for every LUT must set. This task should be done with the SDK.
If LUT values should be retained in the camera after disconnecting the power,
then they must be saved with
UserSetSave.
C.4.2
Full ROI LUT
This section describe the settings for one LUT that is applied to the full ROI.
1.
Set
LUT_EnRegionLUT
(in category
RegionLUT
) to False. This is required to use the full ROI LUT.
2.
Set
LUTEnable
(in category
LUTControl
) to False. This is not mandatory but recommended.
3.
Select LUT 0 by setting
LUTSelector
(in category
LUTControl
) to 0.
4.
Set LUT content as described in Section C.4.4.
5.
Turn on LUT by setting
LUTEnable
to True.
C.4.3
Region LUT
The Region LUT feature is described in Section 11.4.4. Procedure to set the Region LUT:
1.
Set
LUT_EnRegionLUT
(in category
RegionLUT
) to False. This is not mandatory but
recommended.
2.
Set
LUTEnable
(in category
LUTControl
) to False. This is not mandatory but recommended.
3.
Select LUT 0 by setting
LUTSelector
(in category
LUTControl
) to 0.
4.
Set properties LUT_X, LUT_W, LUT_Y and LUT_H (all in category
RegionLUT
) to desired
value.
5.
Set LUT content as described in Section C.4.4.
6.
If two Region LUT are required, then select LUT 1 by setting
LUTSelector
(in category
LUTControl
) to 1 and repeat steps 4 and 5.
7.
Turn on LUT by setting
LUTEnable
to True.
8.
Turn on Region LUT by setting
LUT_EnRegionLUT
(in category
RegionLUT
) to False.
C.4.4
User defined LUT settings
This section describes how to set user defined LUT values. It is assumed that the LUT was
selected as described in Section C.4.2 or Section C.4.3.
For every LUT value the following steps must be done:
1.
Set
LUTIndex
(in category
LUTControl
) to desired value. The
LUTIndex
corresponds to the
grey value of the 12 bit input signal of the LUT.
MAN078 12/2018 V1.1
109 of 111
Summary of Contents for MV0 CMOSIS Series
Page 10: ...1 Preface 10 of 111 MAN078 12 2018 V1 1...
Page 38: ...4 Image Acquisition 38 of 111 MAN078 12 2018 V1 1...
Page 52: ...8 Image Format Control 52 of 111 MAN078 12 2018 V1 1...
Page 78: ...12 Precautions 78 of 111 MAN078 12 2018 V1 1...
Page 88: ...14 Mechanical Considerations 88 of 111 MAN078 12 2018 V1 1...
Page 90: ...15 Troubleshooting 90 of 111 MAN078 12 2018 V1 1...
Page 96: ...18 Support and Repair 96 of 111 MAN078 12 2018 V1 1...