PR-680/680L User Manual
87
Data Code
Description
601
(PMT)
Note
:
To view the exposure time for the last measurement when the
instrument is in Adaptive Exposure mode, send the D13 command.
Output
Format
:
Error Code <qqqqq>,
<Detector> (0=PDA, 1=PMT),
<PrimaryAcc>,
<AddOn1>,
<AddOn2>,
<AddOn3>,
<Aperture>,
<Units> (0=English, 1=Metric),
<Gain> (0=1X, 2=10X, 3=100X, 99=Auto),
<ND Filter> (0=OPEN, 1=ND-1, 2=ND-2, 3=ND-3),
<Max Exposure Time> (in msec),
<Cycles to Average>,
<PMT Filter> (0=None[OPEN], 1=50 KHz, 2=15 KHz, 3=600 Hz, 4=200 Hz,
5=10 Hz)
<Measure Shutter> (0=Closed After Measurement, 1=Always Open)
<Analog Output> (0=Disabled, 1=Enabled)
CRLF
Output Example
:
00000,1,0,-1,-1,-1,2,0,99,0,10000,1,4,0,0
602
(SRM)
Output Format
: Current set report with text labels. For more details, see Report 601
above
Output Example
:
00000,PDA,MS-75,None,None,None,1/4 deg,English,Adaptive,2000 msec,Normal,1 cycle
s,2 deg,No Smart Dark,Standard Sensitivity,No Sync,60.00 Hz,PDA,6000 msec
602
(PMT)
Output Format
: Current set report with text labels. For more details, see Report 601
above
Output Example
:
00000,PMT,MS-75,None,None,None,1/4 deg,English,Adaptive,Open,10000 msec,1 cycles
,200 Hz.,Closed after measurement, Analog Output Disabled