PR-680/680L User Manual
67
R
EMOTE
C
ONTROL
C
OMMAND
S
UMMARY
The following table summarizes all valid Remote Control commands and responses from the
PR-680/680L. Detailed descriptions including parameters passed with each command are detailed in the
Commands
section.
Command
Description
B
Sets LCD backlight level
C
Clears current session instrument errors
D
Downloads data from the PR-680/680L
E
Toggles the Echo (full duplex) mode
F
Measure frequency of light source
I
Requests instrument status or / error report from PR-680/680L
L
Defines measurement title
–
Maximum of 20 characters
.
M
Measure command for the PR-680/680L. Returned datum depends on the
accompanying parameters.
O
Initialize Data Logger Mode. Data Logger Mode acts like a printer, automatically
sending measurement result data over the USB interface following a measurement.
Data types can be defined by the user through switches.
Note: Data Logging is supported on PR-655 / PR-670 firmware versions 2.58 and
above only. Contact Photo Research for upgrade information.
Q
Quit (exit) remote mode.
R
Recall stored measurement
S
Set up measurement parameters
X
Sets LCD contrast level.
TABLE 11 - PR-680/680L REMOTE MODE COMMAND SUMMARY
Usage:
<...>
Optional Entry,
[...]
Mandatory entry
(...)
Comment, NOT PART OF THE COMMAND
Note: The default values, e.g. apertures, exposure time etc. are those used for the measurement
before the instrument was set to Remote Control Mode or, if in Remote Mode, the value from
the previous command.