Command
Description
O
Purpose:
Data Logging is designed to act as an auto print command when making measurements in the stand-
alone (touch screen) operation. It automatically sends requested data types over the USB interface.
Note:
Data Logging is supported on PR-655 / PR-670 / PR-680 Firmware versions 2.58 and above
only. Contact Photo Research for upgrade information.
Once Data Logging is enabled using the following Remote Mode commands, it remains
enabled until cancelled, again using Remote Mode commands, even if the instrument is
turned off.
OS
Status and settings of Data Logger
Syntax
: OS[CR]
Response Code:
00000,Enabled=b,n1,n2,n3...ni[CR][LF]
Where: b has the value 0 or 1 and n1 thru ni are the Data Send Codes stored in the Data Logger Report
List. The Data Logger Report List can hold up to 50 Data Send Codes. See the Measurements and Data
Send Codes section for further details on send codes.
OE
Enable data logging
Syntax
: OE[CR]
Response Code:
00000[CR][LF]
OD
Disable data logging
Syntax
: OD[CR]
Response Code:
00000[CR][LF]
O+n Add Data Report n to Report List (where n = Data Send Code)
Syntax
: O+<data send code>[CR]
Response Code:
00000[CR][LF] If all OK, else NNNNN[CR][LF]
Where: NNNNN = Error code
O-n
Remove Data Report n from Report List (where n = Data Send Code)
Syntax
: O-<data send code>[CR]
Response Code:
00000[CR][LF] If all OK, else NNNNN[CR][LF]
Where: NNNNN = Error code
ORn1,n2,n3,...,ni
Add Data Send Codes n1 thru ni to Report List
Syntax
: ORn1,n2,n3,...,ni[CR]
Response Code:
00000[CR][LF] If all OK, else NNNNN[CR][LF]
Where: NNNNN = Error code
OC
Clear the Report List (remove all Data Send Codes from list)
Syntax
: OC[CR]
Response Code:
00000[CR][LF]
Q
Purpose
: Quit (Exit) Remote mode
Syntax: Q
Response: None