Caution!
For your own safety!
Shutdown in an emergency
is achieved by
pulling out the mains plug from the DC
power supply!
The mains plug must therefore be accessible at all times when the system
is in operation to ensure it can be quickly pulled out of the wall socket in an emergency.
Avoid subjecting the main's connection lead and connection cables to mechanical loads.
Regular visual checks of the mains connection lead should be carried out at short-term
intervals and in particular each time it has been subjected to a mechanical load.
If the mains connection lead of the DC power supply gets damaged, it must be replaced in
order to prevent hazards.
Work to and replacement of the mains connection lead may
only be carried out by specialist personnel holding the qualifications described on
page 3 or by persons who have taken part and successful completed the
corresponding training programmes offered by DewertOkin.
In the Operating Instructions to be drawn up by yourselves, it is essential that you draw the
operator´s attention to the points mentioned here.
3. Getting to Know the System
The
CU155
drive control unit is intended for the German market and complies with the Law
applicable in Germany in implementation of relevant EC Directives.
a)
Product versions
To operate the drive controls,
further components
,
such as from 1 to 4 slave drives, (5
slave drives optional) handset
... are required.
Reverter-Plug
The optionally available reverter-plug (see page 7) allows flexible adjustment of the motor's
running direction to match the application. The controls and handset can therefore be
standardized.
Connectable components
4
Slave Drive
Handset
Voltage Supply
1)
CU155
MEGAMAT
COMFORT DC power supply
1)
MULTITRAXX
1)
MULTIMAT B23
.......
.......
1)
with or without
reverter-plug
5
b) Technical data
Supply voltage..................................................:
24 V DC
Permissible total current input
.......................... :
max. 8.0 A (ID 2min./18min.)
Fuse................................................................. :
T 6.3A - T 8.0 A (depending on version)
1)
Operating mode ............................................ :
intermittent duty ID 2 min./18 min.
Protection classification.................................... :
III
Variations...
......................................... :
v
..............
arious control functions for up to 5 drives
Protection category........................................... :
IP20 (optional IPX4)
Colours............................................................. :
black
Dimensions and weights
Length x width x height of the control system...:
157 x 128.5 x 44 mm
Weight.............................................................. :
approx. 0.3 kg
Ambient conditions for operation, storage and transport
Transport / storage temperature....................... :
from -20 °C to +50 °C
from -4 °F to +122 °F
Operating temperature...................................... :
from +10 °C to +40 °C
from +50 °F to +104 °F
Relative humidity............................................... :
from 30% to 75%
Air pressure.............................................................: from 800 hPa to 1060 hPa
Height......................................................................: < 2000 m
1)
Operating mode =
intermittent duty ID 2 min./18 min.
, i.e. run for 2 min. max. under rated load, then a rest
period of 18 min. must be observed, otherwise operational
failure could occur
!
CU155
03/2020
ID 49623 3.0
03/2020
ID 49623 3.0