4 Testing and Maintenance
Reporting of Test Results
72
Safety (3)
Perform Safety Test (3): Equipment Leakage
Current - Single Fault Condition (Open
Earth)
With mains cable:
Maximum impedance (X2):
<= 300
A)
S(3):P/X2 or
S(3):F/X2
where P=Pass and
F=Fail
Alarming
Functionality
Set the apnea alarm delay to 10 seconds and
then breathe twice into the gas inlet. (Make
sure to set the apnea alarm delay back to its
original setting after completing the test).
Apnea alarm is issued 10
seconds after breathing into the
inlet has stoppped.
A: P or
A: F
where P=Pass and
F=Fail
Test Block Name Test or Inspection to be performed
Expected Test Result
What to Record
on Service
Record