75
5
5
Troubleshooting the Gas
Analyzer
This chapter provides a recommended procedure for locating and identifying faults on the gas analyzer.
It details how to identify hardware problems and how to proceed when measurement related INOPs
occur.
It details how to proceed when errors are flagged for:
• Failed calibration checks and procedures
• Failed diagnostic checks.
Equipment needed for troubleshooting:
Equipment
Philips part # or other recommendation
Accuracy
IntelliVue G1/G5 PC cable
M1013-61005 / 451261005001
n/a
TSI Flowmeter
453564178121
3 ml/min or better
Digital Barometer
Recommended: DRUCK DPI 705, 2bar, absolute
2 mbar or better
Watertrap
M1657B / 989803110871
n/a
Sample Tubing
M1658A / 989803104671
n/a
Calibration Gas Reservoir Bag M1659A / 989803104681
n/a
Calibration Gas
M1662A / 451261001391
n/a
Gas Exhaust Return Line
M1655B / 989803145671
n/a