Installing the Top Mount
2 Installation and Patient Safety
21
N O T E
Make sure the sample gas is routed through the CO
2
absorber before going back to the patient.
1
M1656B Gas Exhaust Return Filter
2
Female luer lock
3
Shorter tube connecting to the ventilation circuit
4
Longer tube connecting to the Anesthetic Gas Exhaust Outlet
Removing the Gas Sample
To
remove
the gas sample from the anesthesia circuit, a scavenging system needs to be connected to the
gas analyzer’s Anesthetic Gas Exhaust. If you intend to use a scavenging system with the gas analyzer,
one of the following parts must also be connected to protect it against malfunction:
1
A ventilator reservoir where the suction pressure does not exceed 70 mbar or
2
A scavenging interface, properly set and maintained (see scavenging interface manufacturer’s
instructions).
N O T E
If you are not returning the gas sample into the patient’s breathing circuit, install the M1655B Exhaust
Return Tubing
without
the M1656B Exhaust Return Filter, shorter tube and the luer lock fitting. See
the Instructions for Use provided with the tubing and filter for further details.
Installing the Top Mount
1
Remove the three rounded head screws on the top of the gas analyzer.