Installing inner attachments
To re-install all the attachments, perform the procedure in reverse order from step 8 on page 24.
Note:
When installing the fan, insert it to the motor
shaft securely. Lightly turn and pull the fan
manually to make sure that it does not touch the
rear panel and is installed securely (Fig. 1).
CAUTION
If the fan is not inserted deep enough, the intended
velocity performance cannot be achieved, and it may
cause culture failure or insufficient decontamination.
Note:
To install the duct, confirm 4 pins are
securely installed in the 4 holes of the duct. (Fig. 2)
CAUTION
If the duct is fixed insecurely, the intended velocity
performance cannot be achieved, which may cause
culture failure or insufficient decontamination.
Note:
When installing the fan cover, position the
long hole of duct with the projection of fan cover
and insert directly (Fig. 3). Same applies for the
humidifying pan cover.
The fan cover may lean by strongly pushing the
head of it in the back. Make sure that there is no
space in the lower fan cover after installing
because the leaning fan cover may have a bad
influence on the camber temperature distribution.
CAUTION
If the fan cover is fixed insecurely, the intended
velocity performance cannot be achieved, which may
cause culture failure.
Note:
Set the tray with only the front edge bent
down (Fig. 4).
1. Position the center hole of the fan with the
projection of the motor shaft.
And insert it deeply.
Fig. 1
Fig. 2
Fig. 4
Fig. 3
1.Position
and insert.
2.Confirm the
direction of fixing
position as shown.
3.Make sure that there is no space in
the lower of fan cover.
2.Lightly turn the fan manually
to make sure that it does not touch the rear panel.
3.Lightly pull the fan manually to make sure that
it is installed.
Bent down