13
LCD touch panel
The following display (called the Top screen) will appear when the power switch is turned ON.
Note:
It takes approximately 20 seconds until Top screen is displayed. During warming-up, “Status: Gas
sensor initializing” is displayed in the Message display field (
13
), and “--.-” is displayed in the Present CO
2
density display field (
4
).
2 6
7 8
3
1
1. Present temperature display field
The current chamber temperature is displayed.
2. Set temperature value display field
The set value of chamber temperature is
displayed. Default setting: 37
o
C.
3. Heating indicator
This lamp lights when the heater is energized.
4. Present CO
2
density display field
The current chamber CO
2
density is displayed.
Nothing is displayed when CO
2
density is set 0 %.
5. Set CO
2
density value display field
The set value of the chamber CO
2
density is
displayed. Default setting: 0 %.
6. CO
2
gas injection indicator
This lamp lights when CO
2
gas is being
injected.
7. CO
2
gas supply line indicator A and B*
1
Current supplying CO
2
gas supply line
(connecting port for CO
2
gas pipe) is
displayed. The connecting port A/B for gas
pipe that is currently supplying CO
2
is
displayed in reverse video and blinks.
8. CO
2
gas supply line select key*
1
This is a key to select CO
2
gas supply line A
or B (Connecting port A or B for CO
2
gas
pipe). When an optional gas auto charger
MCO-21GC is installed, CO
2
gas supply line
A/B changes over automatically when CO
2
gas cylinder is empty. Changeover is also
workable by pressing this key.
*1: Only when an optional component MCO-21GC (Gas auto charger) is installed, this key is workable. They
are not displayed when the MCO-21GC is not installed.
5
4