*1: When an optional electric lock MCO-170EL is installed.
*2: When an optional STD gas auto calibration kit MCO-SG is installed.
Note:
Service key is not available. (Qualified engineer only)
Note:
On the Tools #1 screen, by mistakenly
pressing Temp./Gas Calibration key, Temp./Gas
Calibration screen is displayed.
When this screen is displayed, press Back key to
return to the Tools #1 screen, or press Top key to
return to the Top screen.
When these settings on this screen are changed,
inaccurate temperature or CO
2
density may be
displayed.
●
Operation from other than Menu key
♦
CO
2
gas supply line select key: (Operate) Manual changeover
between CO
2
gas supply line A and B (Page 78)
♦
Message select key:
(Operate) Changing some messages
♦
H
2
O
2
key*
3
: (Operate) H
2
O
2
decontamination (Page 62
~
67)
♦
Unlock key*
4
: (Operate) Unlock the outer door. (Page 68
~
72)
♦
Buzzer key: (Operate) Silencing the buzzer (Alarm is not canceled except for some alarms; page 84)
*3: When H
2
O
2
generator MCO-HP, H
2
O
2
decon board MCO-170HB and electric lock MCO-170EL are
installed in the MCO-170AICUVL.
*4: When an optional electric lock MCO-170EL is installed, the auto lock function is ON.