84
ALARMS, SAFETY, AND SELF-DIAGNOSIS
Alarm or safety
function
Conditions Display
Buzzer
Remote
alarm
Safety
operation
UV lamp failure
The UV lamp burns out.
“Err18: UV Lamp Abnormal” is displayed
in the message display field.
Intermittent
tone
ON
-----
New UV lamp
replacement
The accumulated ON time reaches
approx. 5,000 h.
“Warning: UV Bulb Life” is displayed in
the message display field.
----- ----- -----
Communication
error
When communication between LCD
touch panel and control substrate is
died out or unstable.
“Err56: Communication Failure” is
displayed in the message display field.
----- ----- -----
Warming-up of
gas control
After power switch is turned ON, under
warming-up before temperature is
stable and gas control is enabled.
“Status: Gas sensor initializing” is
displayed in the message display field.
----- ----- -----
●
Table 5
~
7 show the behavior of the alarm (buzzer) and Ring Back function when pressing Buzzer key.
Table 5 In the cases of other than table 6 or table 7.
Remote Alarm setting
Ring Back
setting
Buzzer from CO
2
incubator
Remote Alarm
When pressing
the Buzzer key
When the Ring Back
set time passes
When pressing
the Buzzer key
When the Ring Back
set time passes
ON: Non-interlock
with Buzzer key
ON
OFF
(Alarm is not
canceled)
ON
ON
ON
(Under continuation)
OFF OFF
OFF: Interlock
with Buzzer key
ON ON
OFF (Alarm is
not canceled)
ON
OFF OFF
OFF
Note:
Resolve the cause of the alarm in reference to page 83-85 because the alarm itself is not
deactivated by pressing Buzzer key.
Table 6 In the cases of high limit temperature alarm or Err38 (The outer door opens during H
2
O
2
decontamination; refer to next page).
Remote Alarm setting
Ring Back
setting
Buzzer from CO
2
incubator
Remote Alarm
When pressing
the Buzzer key
When the Ring Back
set time passes
When pressing
the Buzzer key
When the Ring Back
set time passes
ON: Non-interlock
with Buzzer key
ON
ON
ON
(Under continuation)
ON
(Continue)
ON
(Under continuation)
OFF
OFF: Interlock
with Buzzer key
ON
OFF
Note:
Close the outer door when Err38 is activated.
Table 7 In the cases of Err01 (CO
2
gas cylinder empty), Err11, 12 (CO
2
sensor error), Err18 (UV
lamp failure) or door alarm*
2
.
Remote Alarm setting
Ring Back
setting
Buzzer from CO
2
incubator
Remote Alarm setting
When pressing
the Buzzer key
When the Ring Back
set time passes
When pressing
the Buzzer key
When the Ring Back
set time passes
ON: Non-interlock
with Buzzer key
ON
OFF
(Alarm is
canceled)
OFF
(Alarm is already
canceled)
OFF
(Alarm is
canceled*
2
)
OFF
(Alarm is already
canceled*
2
)
OFF
OFF: Interlock
with Buzzer key
ON
OFF
*2: In the door alarm, the remote alarm does not work.
Note:
When Err01 is activated, connect the new CO
2
gas cylinder and press the Buzzer key to stop the
buzzer. In addition, when the optional MCO-21GC is installed and the gas supply is switched to the reserve
gas cylinder, press the Buzzer key and replace the gas cylinder.