24
FFe
ea
attu
urre
e
U
Un
niitt
TMH 071 P
TMH 071 P
TMU 071 P
TMH 071 Y P
TMH 071 Y P
TMU 071 Y P
Connection nominal diameter
Inlet
DN 40 ISO-KF
DN 63 ISO-K
DN 63 CF-F
Outlet
DN 16 ISO-KF/G 1/4“
DN 16 ISO-KF/G 1/4“
DN 16 ISO-KF/G 1/4“
Venting connection
G 1/8”
G 1/8”
G 1/8”
Nominal rotation speed
Hz (1/min)
1500 (90 000)
1500 (90 000)
1500 (90 000)
Standby rotation speed
Hz (1/min)
1000 (60 000)
1000 (60 000)
1000 (60 000)
Start-up time
min
2.5
2.5
2.5
Noise level
dB (A)
< 45
< 45
< 45
Final pressure, backing pump
mbar
< 10
< 10
< 10
Integral leak rate
mbar l/s
< 2 · 10 8
< 2 · 10 8
< 2 · 10 8
Maximum permissible
rotor
temperature
°C 80 80 80
Permissible heat radiation power
W
4
4
4
Volume flow rate for
Nitrogen N2
l/s 33 59 59
Helium
He
l/s 39 50 50
Hydrogen H2
l/s 32 42 42
Compression ratio for
Nitrogen N2
> 10 11
> 10 11
> 10 11
Helium He
6 · 10 6
6 · 10 6
6 · 10 6
Hydrogen H2
> 10 5
> 10 5
> 10 5
Maximum fore-vacuum pressure for
Nitrogen N2
mbar
18 18 18
Helium
He
mbar
18 18 18
Hydrogen H2
mbar 12 12
12
Maximum gas throughput at nominal rotation speed 2)
With water cooling
Nitrogen N2
mbar
l/s
1.3 1.3 1.3
Helium He
mbar l/s
2.7
2.7
2.7
With air cooling 3)
Nitrogen N2
mbar l/s
0.58
0.58
0.58
Maximum gas throughput at intake pressure of 0.1 mbar 4)
With water cooling
Nitrogen N2
mbar
l/s
2 2 2
Helium He
mbar l/s
3
3
3
Hydrogen H2
mbar
l/s
2 2 2
Vertex power characteristics line 5)
A
W / Hz
80 / 1500
80 / 1500
80 / 1500
B
W / Hz
80 / 1500
80 / 1500
80 / 1500
C
W / Hz
60 / 1500
60 / 1500
60 / 1500
D
W / Hz
80 / 1300
80 / 1300
80 / 1300
Final pressure 6)
With rotary vane pumps
mbar
< 1 · 10 -7
<.1 · 10 -7
< 5 · 10 -10
With diaphragm pumps
mbar
< 1 · 10 -7
< 1 · 10 -7
< 1 · 10 -8
Lubricant
TL 011
TL 011
TL 011
Maximum cooling water
consumption with water at 15 °C 6)
l/h 100
100
100
Cooling water temperature
°C
5 - 25
5 - 25
5 - 25
Permissible ambient temperature
with air cooling
°C
0 - 40
0 - 40
0 - 40
Heating power consumption
W
32
32
32
Weight
kg 2.8 3.8 3.8
Permissible magnetic field
mT
3
3
3
Operating voltage
VDC
24 ± 5%
24 ± 5%
24 ± 5%
Duration 7) - / max. current consumption
A
4.1 / 4.6
4.1 / 4.6
4.1 / 4.6
Duration 7) - / max. power
W
100 / 110
100 / 110
100 / 110
Fuse, internal
V
T8A/250
T8A/250
T8A/250
Protection class 8)
IP 30
IP 30
IP 30
Relative humidity
%
5-85 non-condensing
5-85 non-condensing
5-85 non-condensing
9. Technical Data
1)Measured with 20 % helium concentration and a measurement period of 10s
2) Until frequency fall-off; higher gas throughputs with reduced rotation speed.
3) Until ambient temperature 40 °C.
4) Rotation speed of pump may drop below the nominal rotation speed.
5) For gas characteristic lines please refer to section 4.4.
6) In accordance with German Industrial Standard DIN 28428
7) At maximum gas throughput.
8) Protection class IP 54 is afforded for the TC 600 by retro-fitting a cover plate (Accessory)