18
5. Monitoring Operations
5.1. Operations Display Via LED
Certain operations modes of the turbopump and the TC 600
can be ascertained via the two integrated LEDs located on
the front panel of the TC 600.
The following operations modes are displayed:
5.2. Turbopump Temperature
Management
Where impermissible motor temperatures are involved or the
temperature of the casing is too high, the motor current is
reduced.
This can lead to dipping below the set rotation speed switch-
point and results in the turbomolecular pump being switched
off.
LED on the TC 600 glows red: Collective malfunction.
LLE
ED
D
C
Ca
au
usse
e
g
grre
ee
en
n
rre
ed
d
Glows
– power supply OK
– Function ”pumping stati on ON”
carried out
Flashes short
– power supply OK
(1/12s active)
– Pumping station OFF
Blinks
– Mains power supply failure
(1/2s active)
Glows – Collective
malfunction
(for example, run-up time
error, over-temperature,
turbopump or TC 600)
– Switching output 2 active
(low)
Blinks
– Warning
(1/2s active)
(e.g. supply voltage short
circuit to earth, mains
power failure)
Differentiated malfunction and warning signals
are only possible with the use of a DCU, HPU
or PC.
☞