14
TC 600
TC 750
X4
1
1
2
2
3
3
4
n.c.
Mains input
power supply
(90 - 132 /
185 - 265) VAC
TPS XXX/DCU XXX
X2
Protection conducter (PE)
Supply voltage, TC
GND
PE
GY/YE
Br
Bl
1
2
3
4
5
6
FAN
Fan
M
Relay box
FV PUMP
1
2
3
4
5
6
L
N
PE
PE
N
L'
Pumping station (Imax = 16 A)
Mains input, pumping station
115/208/230 VAC
Connection, pumping station
VENT
1
2
3
4
5
6
Venting valve
TVF 005
HEAT/TMS
1
2
3
4
5
6
Relay box
Heating
Mains input, heating
115/208/230 VAC
HEATING
L
N
PE
PE
N
L'
REMOTE
1
2
3
4
5
6
+24 VDC*/max. 50 mA
Venting release
Motor TMP
Pumping station
Heating / Reset (opt. sealing gas)
Standby
PWM on (touch relation 25 - 75 %)
Switching output 1 (24 VDC / max. 50 mA)
Switching output 2 (24 VDC / max. 50 mA)
GND*
Contact current:
max. 6 mA / contact
7
8
9
10
12
13
14
15
11
Contact output 1: Switch point attaind
Contact output 2: Collective malfunction
Analog output 0-10 VDC
RS 485
1
2
3
4
5
6
7
8
1
2
3
4
5
6
7
8
n.c.
+24 VDC* / max. 200 mA (supply voltage, DCU)
n.c.
n.c.
n.c.
RS 485+/ (DO/RI)
GND* (mass connection, DCU)
RS 485- / (DO/RI)
3.13.
Connections Diagram