3.8. Connecting The Electronic Drive Unit
TC 600
The turbopump and the electronic drive unit
TC 600 are connected and together form a
single unit. The connecting cable 8a has to be
ordered separately (see ”Accessories”).
If the turbopump is operated with a display and
control unit DCU 001/DCU100 or the HPU 001,
the remote plug 8d must be disconnected.
➡
Unscrew screw with tooth lock 8c from the TC 600 (above
the connection X4).
1)
➡
Plug the plug X4 on connecting cable 8a into the
connection X4 on the TC 600 and screw in screw 8b.
➡
Secure plug X4 on the TC 600 with a screw and
tooth lock 8c.
1)
➡
Connect plug X2 on connecting cable 8a with power
supply TPS 200/DCU 200 (Accessories) on connection X2.
➡
Using screws and toothlock 8c (included with the cable
consigment) secure plug X2 to power supply 105.
1)
1)
Only with cable PM 051 843 -T (see section “Accessories”)
☞
PLEASE NOTE
☞
PLEASE NOTE
X4
8
8b
X4
8c
8a
8c
X2
X1
S1
X2
105
8d
Connecting the TC 600 with power supply TPS 100 or DCU 100, mounting diagram for shielded connection cable
X1 Mains connection
X2 Connection, power supply
X4 Connection TC 600
S1 ON/OFF switch
8
Electronic Drive Unit TC 600
8a Connecting cable, TC 600 – TPS/DCU
8b Screw
8c Screw with tooth lock (3 pieces)
8d Remote plug
105 power supply
Once operating voltage has been supplied, the
TC 600 performs a self test on the supply
voltage. The supply voltage for turbopumps
TMH 071 / TMU 071 is 24 VDC ±5% in accor-
dance with EN 60 742.
3.9. Installing The Power Supply
Voltage may only be supplied with the Pfeiffer
Vacuum power supplies (accessory). The use of
other power supplies requires the prior agree-
ment of Pfeiffer Vacuum and equalization with
the valid specification (specifications available
on request).
The mains connection must be freely
accessible at all times.
CAUTION
CAUTION
CAUTION