6 Operation
6.1 Commissioning
Important settings and function-related variables are factory-programmed into the vacuum pumps elec-
tronic drive units as parameters. Each parameter has a three-digit number and a description. Parame-
ters can be used via Pfeiffer Vacuum display and control panels or externally via RS-485, using the
Pfeiffer Vacuum protocol.
For both parameter sets, the DCU acts as the user interface. HiCube Eco controls two independent pa-
rameter sets, for the turbopump and backing pump. The backing pump parameters are shown hierarchi-
cally in the display, and marked with a symbol.
Parameter
Name
Designation
Option
HiCube pumping station
30 Eco
80 Eco
300 Eco
[P:001]
Heating
Heating
no heating
not available
0
0
with heating
1
1
[P:012]
EnableVent Enable vents
1
1
1
[P:019]
Cfg DO2
Output DO2 config-
uration
22 = Backing
pump standby
22 = Backing
pump standby
22 = Backing
pump standby
[P:023]
MotorPump Motor pump
1
1
1
[P:024]
Cfg DO1
Output DO1 config-
uration
15 = Pump state
15 = Pump
state
15 = Pump state
[P:025]
OpMode
Backing pump op-
erating mode
1 = Interval
mode
1 = Interval
mode
1 = Interval
mode
[P:035]
Cfg Acc A1
Connection A1 ac-
cessories configu-
ration
Air cooling
4 = Fan (tem-
perature control-
led)
4 = Fan (tem-
perature con-
trolled)
4 = Fan (temper-
ature controlled)
Water cool-
ing
2 = heating
2 = heating
2 = heating
[P:036]
Cfg Acc B1
Connection B1 ac-
cessories configu-
ration
1 = Venting
valve, normally
closed
1 = Venting
valve, normally
closed
1 = Venting
valve, normally
closed
[P:708]
PwrSVal
Default power con-
sumption
for
MVP 015-2
90 %
90 %
90 %
for
MVP 015-4
100 %
100 %
100 %
for MVP 030
100 %
100 %
100 %
[P:710]
Swoff BKP
Backing pump off
threshold in interval
mode
25 W
25 W
45 W | MVP 015
52 W | MVP 030
[P:711]
SwOn BKP
Backing pump on
threshold in interval
mode
40 W
40 W
55 W | MVP 015
62 W | MVP 030
[P:794]
Param set
Parameter set
1 = advanced
1 = advanced
1 = advanced
[P:795]
Servicelin
Service row view
309 = Actual
speed
309 = Actual
speed
309 = Actual
speed
[P: 030]
ValveMode
Gas ballast configu-
ration
for MVP 015
0 = auto
0 = auto
0 = auto
[P: 717]
StdbySVal
Default rotation
speed in standby
for MVP 015
66 %
66 %
66 %
Operation