1
Fig. 15:
Backing pump transport lock
1
Transport lock hexagon nut
Required tool
● Hexagonal wrench (13 WAF)
Procedure
1. Rotate the transport lock hexagon nut upwards to the stop.
2. Also release the transport lock on the opposite side.
5.9 Grounding the pumping station
Fig. 16:
Example: Connecting the grounding cable
Required aids
● Screw M4 × 8
● Shake proof washer M4, if required
● Suitable grounding cable with M4 cable lug
Procedure
1. Use a suitable grounding cable to divert applicative interferences.
2. Route the connection in accordance with locally applicable provisions.
3. Use the turbopump ground terminal (M4 female thread).
5.10 Establishing the mains connection
WARNING
Danger to life from electric shock due to improperly performed installation
The device uses voltage that is dangerous on contact as the electrical power supply. Potentially fatal
situations arise due to unsafe or incorrectly installation when reaching into the device.
► Ensure that the system is safely integrated into an emergency off safety circuit.
► Do not carry out any unauthorized modifications or changes to the device.
Installation
30/60