Maintenance
99
Lamp Energy Attenuator
An attenuator is located in the lamp compartment between the deuterium lamp and
the beam entrance slit.
If required you can decrease or increase the energy by placing the attenuator into, or
taking the attenuator out of, the beam.
The attenuator is set at the factory, normally in the out position.
WARNING
Electrical Hazard
Switch off the spectrometer and remove the plug from the electrical
supply before starting with the replacement.
Risk of Burns
If the old lamp was lighted: allow it to cool before proceeding with the
replacement.
UV Radiation
The lamps emit intense radiation which can damage your eyes.
Do not open the lamp compartment when the lamps are on.
Do not gaze into a lighted lamp.
Operate the attenuator as follows:
1. Open the lamp compartment.
2. Remove the lamp baffle by slackening the thumbscrews and lifting the lamp
baffle vertically upward.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Summary of Contents for Lambda 25
Page 9: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 23: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 37: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 55: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 73: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 87: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 107: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 113: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...