Optics
Beam center height
15 mm above cell holder bottom
Beam cross-section
1 nm slit ca. 0.6 mm x 9 mm (width x height)
at focal point of sample and reference beam in
sample compartment
Optical pathlength in
sample compartment
121 mm
Grating (Monochromator)
Holographic concave grating with 1053 lines/mm
in the center
Radiation sources
Prealigned deuterium and halogen lamps
Detector
Photodiodes
(One for the sample beam and one for the
reference beam)
Abscissa
Wavelength range
190 nm to 1100 nm;
0 nm for alignment purposes
Wavelength accuracy
±0.1 nm
Wavelength reproducibility
±0.1 nm
Spectral bandwidth
1 nm (fixed slit)
Lamp change
Automatically at 326 nm
(software selectable over the whole
wavelength range)
Scan speeds
7.5, 15, 30 , 60, 120, 240, 480, 960,
1920 and 2880 nm/min
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Summary of Contents for Lambda 25
Page 9: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 23: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 37: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 55: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 73: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 87: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 107: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 113: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...