98
Conducted RF
IEC61000-4-6:2009
Radiated RF
IEC61000-4-3:2006
3 Vrms
150k
H
Z
~ 80M
H
Z
3 V/m
80M
H
Z
~ 2.5G
H
Z
3 Vrms
3 V/m
Portable and mobile RF communications equipment
should not be used near any part of this equipment,
including cables, that is closer than the recommended
separation distance calculated from the equation for
transmitter frequency.
Recommended separation distance:
d = 1.2 √P
d = 1.2 √P80M
H
Z
~ 800M
H
Z
d = 1.2 √P800M
H
Z
~ 2.5G
H
Z
Remarks : UT is the a.c main voltage prior to applying the test level.
*Remark 1)
Input/output Line test is not applicable because there is no input/output line.
*Remark 2)
Tests between line and ground are not applicable since there is no connection terminal.
Instructions and manufacturer declarations - Electromagnetic immunity.
This unit is intended for use in the electromagnetic environment specified below. In addition, the user should
use this product in the following environment.
Immunity Test
Compatibility Level
IEC60601 Test Level
Electromagnetic environment - Guideline