97
Surge
IEC61000-4-5:2005
Power frequency
(50/60H
Z
)
magnetic field
IEC61000-4-8:2009
<5% UT
(95% drop from UT)
0.5 Cycle
70% UT
(30% drop from UT)
25 Cycle
<5% UT
(95% drop from UT)
300 Cycle
3 A/m
3 A/m
40% UT
(60% drop from UT)
5 Cycle
<5% UT
(95% drop from UT)
0.5 Cycle
40% UT
(60% drop from UT)
5 Cycle
70% UT
(30% drop from UT)
25 Cycle
<5% UT
(95% drop from UT)
300 Cycle
The mains supply should be of the quality of a
commercial or hospital environment.
If the user needs to continue to use during
a power outage, it is recommended that the
unit be powered by an uninterruptible power
supply or a battery.
The power frequency magnetic field should have
the general position level characteristics of a
typical commercial or hospital environment.
Immunity Test
Compatibility Level
IEC60601 Test Level
Electromagnetic environment - Guideline