96
Electrostatic
discharge (ESD)
IEC61000-4-2:2008
EFT Burst
IEC61000-4-4:2012
Surge
IEC61000-4-5:2005
Air: ± 8 kV
Contact: ± 6KV
HCP/VCP: ± 6KV
Power Supply Line:
± 2KV (*Remark 1)
Line-Line: ± 0.5,
1KV (*Remark 2)
Air: ± 8 kV
Contact: ± 6KV
HCP/VCP: ± 6KV
Power Supply Line: ±2KV
Input-Ouput Line: ±1KV
Line-Line: ± 1KV
Line-Ground: ± 2KV
The floor must be wood, concrete or
ceramic tile. If the flooring is synthetic, the
relative humidity should be at least 30%.
The floor must be wood, concrete or
ceramic tile. If the flooring is synthetic, the
relative humidity should be at least 30%.
The mains supply should be of the quality
of a commercial or hospital environment.
Immunity Test
Compatibility Level
IEC60601 Test Level
Electromagnetic environment - Guideline
Instructions and manufacturer declarations - Electromagnetic immunity: This unit is intended for use in the
electromagnetic environment specified below. In addition, the user should use this product in the following
environment.