Continued from previous page
Viewing data in
Microsoft® Excel
The table below is an example of temperature data from the warmer.
Result
The temperature data has now been viewed.
Time Stamp
Chamber Mode
Power
Door
Fan
Temperature Set-point
Errors
5/1/2019 8:52
bottom
off
open
off
71.2
104
Diagnostics.SensorSho
rtErrors=0x0c
5/1/2019 8:57
bottom
off
open
off
70.9
104
Diagnostics.SensorSho
rtErrors=0x0c
5/1/2019 9:01
bottom
INJ
on
closed
on
69.9
104
Diagnostics.SensorSho
rtErrors=0x0c
5/1/2019 9:02
bottom
INJ
on
closed
on
69.3
104
Diagnostics.SensorSho
rtErrors=0x0c
5/1/2019 9:07
bottom
INJ
on
closed
on
79.3
104
Diagnostics.SensorSho
rtErrors=0x0c
5/1/2019 9:12
bottom
INJ
on
closed
on
91.2
104
Diagnostics.SensorSho
rtErrors=0x0c
5/1/2019 9:17
bottom
INJ
on
closed
on
98.9
104
Diagnostics.SensorSho
rtErrors=0x0c
5/1/2019 9:22
bottom
INJ
on
closed
on
101.8
104
Diagnostics.SensorSho
rtErrors=0x0c
5/1/2019 9:27
bottom
INJ
on
closed
on
104
104
Diagnostics.SensorSho