O
PERATION
Combination Warmer
▪
Operator’s Manual
▪
M N-47424
▪
Rev 1
▪
4/21
34
Continued from previous page
2.
Set
the temperature using the arrow buttons
.
3.
Open
the door and load the fluid or blankets into the warmer.
Close
the door.
During the warming
process
4.
Press
the temperature recall button
to view the measured cavity
temperature.
The measured cavity temperature will display for five seconds. Then, the
set-point temperature will display.
Result
The fluids or blankets are now warming.
NOTE:
The irrigation temperature set-point range is
90°F – 150°F (32°C – 66°C).
The injection temperature set-point range is 90°F – 104°F
(32°C – 40°C).
The blanket temperature set-point range is 90°F – 160°F
(32°C – 71°C).